Specifications
|
Power switch |
220V/6A |
|
Motor power |
750W |
|
Maximum capacity |
100 liters / 200 liters |
|
Usage voltage |
220V/50Hz |
|
Motor capacitance |
10μF / 450V |
|
Operating temperature |
<5℃ & < 50℃ |
Methods of use
Washing
(1) Remove the plastic sealing film of the filter core and install the filter.
(2) Close the Product valve and drain valve. Open the mixing valve and add water. Turn on the pump switch and start cleaning when the water reaches the desired amount.
(3) Open the spray valve in the middle for spray operation. Turn off the pump switch after 5 minutes.
(4) Open the drain valve. When there is no liquid in the bucket. The mixing bucket is wash over. (If you use disinfectant to clean, please remove the filter element in advance)
Mixing
(1) Close the product valve and drain valve. Open the mixing valve, Turn on the water inlet, add water to 90% of final level of water, turn on the pump switch and start feed the powder.
(2) When the water product at the bottom of the bucket can be clearly seen (about 30min), turn off the pump switch.
(3) Turn on the water inlet switch and replenish the water to the final liquid level. Turn on the pump switch for 1 minute.
Filling
Place the small bucket at the product of the finished product and open the product valve. The Dialysis concentrate flows out of the product through the filter, and the mix valve is adjusted to control the flow. Turn off the pump switch when all dialysis concentrate flows out.
Draining
Open the drain valve to drain the residual liquid.
Precautions
- The product is only suitable for the preparation of the dialysis concentrated solution, and is not used as a storage container for the dialysis concentrated solution.
- In the use of the product, to ensure that there is a reliable grounding wire, to ensure the normal work of the water pump safety.
- Before and after each preparation of the concentrated dialysis solution, the mixing tank must be cleaned to ensure that there is no residual liquid in the tank
- It is prohibited to use this machine to prepare other solutions.
- The mixing bucket should be sterilized on a regular basis. The first use must be sterilized.
- This type of mixer uses a 10 inch 5 μ m melt spray filter core.
- The filter core is recommended to be replaced once a month. If you find that the filter core is obviously dirty, please replace it immediately.
Strictly prohibited to run the mixer without water!
Disinfection and Cleaning
The materials use PP, UPVC and other corrosion resistant materials in this product. It is suitable for peracetic acid as disinfectant. If users choose other disinfection methods, it is necessary to pay attention to whether the selected disinfection method is suitable for the product material to avoid unnecessary damage. If in doubt, please consult us.
Please follow below process:
1. Prepare 2.5 liters of peracetic acid disinfectant solution (15-17%), Wear a thickened cotton mask and gloves. Some kinds of peracetic acids need to be mixed for 24 hours before use
2. First remove the filter element of the filter, and then tighten the filter shell.
3. Open the discharge valve and close the mixing tank after emptying thoroughly.
4. Open the discharge valve, and empty the mixing tank completely, and then close the mixing tank. Pour the disinfectant into the mixing tank, and fill the water till half of the maximum capacity of the mixing tank. Take 200L mixing tank as an example, fill with 100L water
5. Put the lid on and turn on the pump for 20 minutes.
6. During this period, open the valve of the liquid intake port, so that a part of the liquid containing disinfection components is discharged. Close the liquid withdrawing valve. Open the spray valve for 3 minutes and reset again.
7. Turn off the pump and stand for 20 minutes.
8. Open the discharge valve and the liquid withdrawing valve of the mixing machine to make it drain and clean.
9. Fill with water to ⅔ of the mixing machine's maximum capacity, open the lid and turn on the mixing pump for 5 minutes. When the liquid is out and then pour it back to the mixing machine during opening the liquid port valve. Open the spray valve for 3 minutes and then reset it.
10. Open the discharge valve and the liquid withdrawing valve of the mixing machine, and discharge it clean.
11. Repeat Steps 9 and Steps 10 and test for disinfectant residue at discharge. If the disinfection residue is qualified, the disinfection shall be terminated. If the residue does not pass, repeat Steps 9 and Steps 10 until the discharges are cleared.
12. The residue test is qualified, and the new filter element is loaded into the filter. Then tighten the filter housing. The frequency of disinfection of the stirred tank will be adjusted according to the results of the bacteria test of the concentrated solution prepared. Generally, Mixing for solution A does not support bacterial growth.
Rinse the mixing bucket: fill water to more scale than usual, and turn on the mixing pump and circulate for 10 minutes, and then drain clean.